Tensys Medical, Inc.
       


Primary Indication

The T-Line® is indicated for measuring arterial blood pressure and pulse rate in adults who weigh between 90 and 400 lb. and who are between 4'6" and 6'6" tall.

Contraindications

The T-Line is contraindicated in patients with:

  • No palpable radial artery pulse
  • Anatomical abnormalities (skin grafts, cysts, cellulitis) or injuries (scarring, cuts, burns, bruising) at the sensor site
  • An intravenous line placed at the sensor application site or within the medial/lateral travel range of the sensor
  • An AV shunt in the T-Line® arm
  • A known sensitivity to pressure stimuli (dermatographism)

Features and Specifications

Measurement Range

Systolic Pressure: 30-245 mm Hg
Diastolic Pressure: 20-220 mm Hg
Pulse Rate: 30-250 BPM (beats per minute)

Electrical

Power Requirements

Input: 100-240 VAC, 50/60 Hz, 0.35 Amperes max
Fuse Rating: 5x20 mm, 0.4 Amperes
Instrument Class: IEC Class II, Conforms to IEC 60601-1

Physical

Weight: 3.75 lbs. (excludes bracelet)
Size: 7" (w) x 8.5" (h) x 3" (d) (excludes pole clamp and bracelet)
Display: TFT-LCD, Color, 1/4 VGA

Environmental

Operating Temperature: +15°C to +40°C
Storage Temperature: -10°C to +60°C
Operating Humidity: 15% to 90% non-condensing
Storage Humidity: 15% to 90% non-condensing

Accuracy

Systolic Pressure: Mean difference is within ± 5 mm Hg; standard deviation not exceeding 8 mm Hg*
Diastolic Pressure: Mean difference is within ± 5 mm Hg; standard deviation not exceeding 8 mm Hg*
Mean Pressure: Mean difference is within ± 5 mm Hg; standard deviation not exceeding 8 mm Hg*
Pulse Rate: ± 3 BPM

T-Line® Radial Artery Sensor

  • Single patient use
  • Provided non-sterile, 20 units/case
  • Maximum duration of use: 18 hrs
  • Latex-free

*Accuracy data collected and reported per AAMI SP-10:2002 Accuracy Specification 4.4.5.2.B (Intra-arterial Method as a Reference Standard) for Non-invasive Blood Pressure Monitors.

View Product Comparison
 