Optimum safety monitoring in the OR requires early enough warning of
hemodynamic changes to allow intervention before significant consequences
develop. The TL-150 accomplishes this and more:
Non-invasive beat-to-beat monitoring
Meets AAMI standards* without the invasiveness of an A-Line, or the
delayed readings of an intermittent oscillometric cuff.
Validated accuracy
Clinically proven to provide timely, accurate measurements and a real-time
waveform equivalent to the accuracy of an arterial cannula.
Early warning
Enables rapid detection of hemodynamic changes that may signal serious
complications.
Early response
Allows immediate intervention for unanticipated events.
Superior control
Enables routine monitoring of even the most difficult patients, and provides
immediate feedback on the impact of interventions for optimal blood
pressure management.
Highly robust
Tolerates significant patient movement.
Easy to learn and use
Hands-on setup time is under 2 minutes.
The TL-150 has been validated against a contralaterally placed invasive A-Line
reference in a statistically significant surgical patient group in accordance with
AAMI SP-10:2002.* Nearly 150,000 paired data points over more than 57 hours of monitoring were collected from this
surgical population and analyzed. The resultant mean and standard deviations
were computed for the three blood pressure measurements, as summarized
below:
Systolic Pressure
Diastolic Pressure
Mean Pressure
Mean Error
1.0 mm Hg
0.8 mm Hg
1.3 mm Hg
Standard Deviation
7.2 mm Hg
7.5 mm Hg
6.2 mm Hg
These data show that the TL-150 is well within the AAMI SP-10:2002 Accuracy Specification of a mean error not exceeding ±5 mm Hg and a standard deviation not exceeding 8 mm Hg.*
Case history:
A 61-year-old female
Revascularization history:
Colitis, PVD, and extensive smoking
Comments:
Unanticipated rapid increase in blood pressure
*Accuracy data collected and reported per AAMI SP-10:2002
Accuracy Specification 4.4.5.2.B (Intra-arterial Method as a Reference Standard)
for Non-invasive Blood Pressure Monitors.